QC-200-S-005-1706E
AUTOCLAVE
(Steam Sterilizer)
CL- 32S / 32L
40S / 40M / 40L
Instruction Manual
Make sure to read this manual to use this product safely.
Place this manual in a convenient location for easy reference.

Questions and Answers

  • K
    Kurt HubbardSep 10, 2025
    Why does evacuating residual air not stop at 100C in my ALP CL- 40S?
    • M
      martinezcolleenSep 10, 2025
      If the ALP Laboratory Equipment does not stop evacuating residual air around 100? and the temperature doesn't rise, it could be due to a breakage of the exhaust valve or because the autoclave is installed over 1000m above sea level. Contact your local distributor, because this autoclave cannot be used over 1000m above sea level.
  • R
    robertssarahAug 23, 2025
    What to do if the lid lock lever on my ALP CL- 40S cannot be released?
    • J
      Julie RamirezAug 23, 2025
      If the lid lock lever of the ALP Laboratory Equipment cannot be released, ensure the power switch (breaker) is turned on. If it is on, and the TEMPERATURE DISPLAY indicates above 80° after operation complete or the COMPLETE LAMP lights up after operation, press the STOP KEY to release the lid lock.
  • S
    Stephen PetersenAug 25, 2025
    Why does the breaker trip immediately when I turn on my ALP CL- 40S Laboratory Equipment?
    • G
      Gina WestAug 25, 2025
      If the breaker trips immediately after powering on the ALP Laboratory Equipment, it could be due to electrical leakage (possibly from insulation deterioration or water leakage at the heating element) or a short circuit (in the heating element or elsewhere). In any of these cases, contact your local distributor.
  • K
    Kenneth BrownAug 27, 2025
    Why is the safety valve activating on my ALP CL- 40S?
    • M
      Margaret RandallAug 27, 2025
      If the safety valve activates on the ALP Laboratory Equipment (steam blowing from the bottom plate), it may be because residual air is not evacuated enough, causing abnormal pressure, or the chamber pressure surpassed set pressure of safety valve due to the malfunction of temperature sensor. Ensure that the quantity of objects inside the chamber is less than 60% of its capacity, with adequate space between them. If the issue persists, contact your local distributor.
  • T
    Tyler OlsonAug 29, 2025
    Why isn't sterilization completing in my ALP CL- 40S Laboratory Equipment even at the set temperature?
    • V
      vjensenAug 29, 2025
      If the ALP Laboratory Equipment reaches the set temperature but sterilization isn't completed because the chamber doesn't fill with saturated steam, it could be due to several factors: * Too much capacity of objects in chamber. Ensure that the quantity of objects inside the chamber is less than 60% of its capacity, with adequate space between them. * Shortage of sterilization time. Increase sterilization time appropriately, especially for liquid objects or large quantities. * Residual air inside container prevents penetration of steam. If using disposable plastic bags, ensure they are filled to less than 60% capacity and have an open mouth. * Air evacuation is insufficient due to bent exhaust hose. Make sure the exhaust hose is not bent.
  • J
    jennifer71Nov 25, 2025
    What does Er1 mean on my ALP CL- 40S and how do I fix it?
    • D
      Debra WilliamsonNov 25, 2025
      If 'Er1' appears on the ALP Laboratory Equipment display with an alarm and the operation stops, it could be due to several reasons: the autoclave is working without water in the chamber, the drain valve is open, there is breakage of water lack preventive device or over pressure detector, over pressure detector activated, residual air in the capacious container or disposable plastic bag swelled or air evacuation is insufficient due to bent exhaust hose. To address this, turn off the power switch and let the autoclave sit for at least one hour to cool down, then pour water into the chamber and close the drain valve. In case of objects being put in a disposable plastic bag, quantity of objects to be put inside disposable plastic bag shall be less than 60% of capacity of the chamber and disp...
  • N
    Nancy NelsonSep 18, 2025
    Why does drainage not stop from my ALP CL- 40S Laboratory Equipment?
    • D
      Dana RileySep 18, 2025
      If drainage from the ALP Laboratory Equipment doesn't stop, it suggests a breakage of the drain valve, possibly due to glass or other objects damaging the valve's Teflon ball. Contact your local distributor.
  • A
    Ashley CampbellSep 22, 2025
    What should I do if the chamber temperature of my ALP Laboratory Equipment doesn't rise?
    • B
      bentleyjefferySep 22, 2025
      If the chamber temperature of the ALP Laboratory Equipment does not rise after pressing the START KEY, it could be due to breakage of the heating element or burning of the wire connected to the heating element. Contact your local distributor.
  • B
    bernarderikaSep 8, 2025
    What does error code Er3 mean on my ALP CL- 40S?
    • R
      Robert StarkSep 9, 2025
      If 'Er3' is shown on the ALP Laboratory Equipment display with an alarm, it indicates that the chamber temperature rose 2? higher than the setting temperature during sterilization. Ensure the distance between the chamber wall and objects is more than 10mm. Oleaginous objects must not be sterilized and if the problem persists, contact your local distributor.
  • D
    Dr. Roberto BellSep 5, 2025
    Why is there steam leaking from the lid of my ALP CL- 40S Laboratory Equipment?
    • J
      josephsmithSep 5, 2025
      If there's steam leaking from the lid of the ALP Laboratory Equipment, it could be due to deterioration of the lid gasket (cracks or hardening) or misalignment of the lid and chamber. Contact your local distributor to resolve this issue.

Summary

Safety Guidelines

Operational Safety

Usage and Cleaning Safety

Essential safety guidelines for operating and cleaning the autoclave.

Installation Safety

Critical safety advice for installing the autoclave correctly.

Operation Safety Notes

Critical Operation Warnings

Key warnings and cautions to prevent accidents during autoclave operation.

Power Failure Handling

Power Failure Safety

Safety measures and procedures to follow during a power failure.

Operation Panel Guide

Installation and Transportation

Transportation Safety

Safety guidelines and precautions for moving the autoclave safely.

Installation Location Requirements

Installation Environment and Safety

Installation Warnings and Cautions

Critical warnings and cautions regarding autoclave installation setup.

Electrical Connections and Grounding

Power Cable and Grounding Safety

Safe procedures for connecting the power cable and ensuring proper grounding.

Exhaust Cooling Tank Safety

Filling and Usage Cautions

Crucial warnings and cautions for safely filling and using the exhaust cooling tank.

Basic Operation Procedure

Pre-Operation Procedures

Pre-Operation Warnings and Checks

Essential warnings, checks, and procedures before starting autoclave operation.

Water Supply Procedure

Water Supply Safety and Cautions

Safety guidelines and cautions for supplying water to the chamber.

Loading and Exhaust Procedures

Object Loading Guidelines

Loading Capacity and Safety

Rules for loading object quantity, accessory use, and lid contact safety.

Disposable Bag Usage Safety

Disposable Bag Warnings

Critical safety warnings for using disposable plastic bags in the autoclave.

Lid Closure and STOP Key Operations

Lid Closure, Locking, and Safety

Procedure and safety for closing and locking the autoclave lid securely.

Operation Pattern Settings

Temperature and Time Configuration

Exhaust Settings and Safety

Starting the Operation

START Key Operation and Safety

Instructions for initiating the sterilization cycle using the START key with safety.

Operation Completion Sequence

Operation Completion Safety

Post-Operation Warnings and Cautions

Critical warnings and cautions after the autoclave operation has finished.

Unloading and Shutdown Procedures

Unloading Objects and Lid Safety

Safe procedures and precautions for unloading sterilized items and lid handling.

Stopping the Autoclave Safely

Stopping Procedure and Safety

Safety guidelines for stopping the autoclave and releasing the lid lock.

Ensuring Sterilization Reliability

Safety Devices Overview

Safety Device Warnings and Cautions

General safety warnings and cautions related to the autoclave's safety devices.

Troubleshooting Guide

Troubleshooting Precautions and Error Codes

Safety advice, precautions, and a reference list of autoclave error codes.

Troubleshooting Common Issues

Troubleshooting Advanced Issues

Troubleshooting Error Codes (Continued)

Maintenance Schedule

Daily Maintenance Checks

Technical Specifications

Disposal and Dismantlement

Disposal Safety Warning

Safety warning regarding electrical shock and handling during disposal.

ALP CL- 40S Specifications

General IconGeneral
BrandALP
ModelCL- 40S
CategoryLaboratory Equipment