Notes on the operation
Be sure to close lid to prevent pinching dusts before operation.
If there is something between lid and chamber which prevents lid from closing, the lid may burst at high pressure
which cause fatal accident or serious injury. Then operator may get burned by steam leaked from chamber.
Be sure to slide lid lock lever slowly to close position (left end) before operation.
If lock is not enough, operator can not start operation (Er4 is shown). But if operator can start operation by
malfunction of detection part, lid may explode at high pressure which cause fatal accident or serious injury.
Objects (including disposable plastic bag) must be put in stainless wire basket or stainless bucket
when sterilization.
If operator loads objects into chamber directly, it may cause choking pipes which may lead to serious accident.
Do not open lid forcibly during operation.
Opening lid forcibly during operation may cause fatal accident or serious injury.
Do not open lid until chamber pressure goes down to “0MPa” when operation completed or
operation interrupted.
Steam spouts out of chamber by residual pressure which may cause burn injury.
In case of sterilizing liquid objects, cool down objects enough after sterilization. Then unload
objects slowly.
Liquid objects keep high temperature after sterilization complete since temperature of liquid cool down slower than
temperature in chamber. Therefore extreme temperature change when unloading may cause high temperature liquid
spouting. It may cause burn injury.
In case of sterilizing liquid objects, metallic or heat-resistance glass container must be used.
A glass container without heat-resistance may be easy to break by extreme temperature change and operator may
be injured by this breakage.
In case of sterilizing liquid objects (EX. liquid media), do not close container with lid or bung which
has no breathability.
It may cause rise of pressure in container, because temperature in container cool down slowly. Especially glass
container is easy to explode if it is damaged.
Do not touch autoclave more than necessary.
If operator touches autoclave more than necessary, it may cause serious accident.
Do not open drain valve under following conditions.
- during operation
- chamber temperature is high
- chamber pressure is not 0MPa
It may cause burn injury by blow-out of boiling water and / or steam, water lack, and/or malfunction.
Do not touch power switch(breaker) by wet hand.
Operator may get electric shock.
Use a heat-resistance glove when operator unloads objects after operation complete.
If operator touches chamber and/or objects without heat-resistance glove, operator may get burned.
Do not touch exhaust bottle and exhaust hose during air evacuation and exhaust.
It may cause burn injury. Do not touch them until cooling down enough.
Do not touch stainless lid directly when opening or closing lid.
It may cause burn injury. Do not touch it until cooling down enough.
Genuine accessories and spare parts made by ALP Co., Ltd. must be used.
Using other company’s products may cause accident and /or breakage of autoclave.