QC-200-S-005-1706E
AUTOCLAVE
(Steam Sterilizer)
CL- 32S / 32L
40S / 40M / 40L
Instruction Manual
Make sure to read this manual to use this product safely.
Place this manual in a convenient location for easy reference.

Questions and Answers

  • C
    cohenjustinSep 10, 2025
    Why does evacuating residual air not stop at 100C in my ALP Laboratory Equipment?
    • A
      Amber FrancisSep 10, 2025
      If the ALP Laboratory Equipment does not stop evacuating residual air around 100? and the temperature doesn't rise, it could be due to a breakage of the exhaust valve or because the autoclave is installed over 1000m above sea level. Contact your local distributor, because this autoclave cannot be used over 1000m above sea level.
  • D
    Deborah RogersAug 23, 2025
    What to do if the lid lock lever on my ALP CL- 40L Laboratory Equipment cannot be released?
    • N
      Nicholas ThompsonAug 23, 2025
      If the lid lock lever of the ALP Laboratory Equipment cannot be released, ensure the power switch (breaker) is turned on. If it is on, and the TEMPERATURE DISPLAY indicates above 80° after operation complete or the COMPLETE LAMP lights up after operation, press the STOP KEY to release the lid lock.
  • Y
    ybraunAug 25, 2025
    Why does the breaker trip immediately when I turn on my ALP CL- 40L Laboratory Equipment?
    • J
      James RyanAug 26, 2025
      If the breaker trips immediately after powering on the ALP Laboratory Equipment, it could be due to electrical leakage (possibly from insulation deterioration or water leakage at the heating element) or a short circuit (in the heating element or elsewhere). In any of these cases, contact your local distributor.
  • L
    Laurie BennettAug 27, 2025
    Why is the safety valve activating on my ALP Laboratory Equipment?
    • J
      Julie GarciaAug 27, 2025
      If the safety valve activates on the ALP Laboratory Equipment (steam blowing from the bottom plate), it may be because residual air is not evacuated enough, causing abnormal pressure, or the chamber pressure surpassed set pressure of safety valve due to the malfunction of temperature sensor. Ensure that the quantity of objects inside the chamber is less than 60% of its capacity, with adequate space between them. If the issue persists, contact your local distributor.
  • H
    hammondbrendaAug 29, 2025
    Why isn't sterilization completing in my ALP CL- 40L even at the set temperature?
    • A
      Abigail WilsonAug 29, 2025
      If the ALP Laboratory Equipment reaches the set temperature but sterilization isn't completed because the chamber doesn't fill with saturated steam, it could be due to several factors: * Too much capacity of objects in chamber. Ensure that the quantity of objects inside the chamber is less than 60% of its capacity, with adequate space between them. * Shortage of sterilization time. Increase sterilization time appropriately, especially for liquid objects or large quantities. * Residual air inside container prevents penetration of steam. If using disposable plastic bags, ensure they are filled to less than 60% capacity and have an open mouth. * Air evacuation is insufficient due to bent exhaust hose. Make sure the exhaust hose is not bent.
  • H
    hrobinsonNov 25, 2025
    What does Er1 mean on my ALP Laboratory Equipment and how do I fix it?
    • D
      Daniel FoxNov 25, 2025
      If 'Er1' appears on the ALP Laboratory Equipment display with an alarm and the operation stops, it could be due to several reasons: the autoclave is working without water in the chamber, the drain valve is open, there is breakage of water lack preventive device or over pressure detector, over pressure detector activated, residual air in the capacious container or disposable plastic bag swelled or air evacuation is insufficient due to bent exhaust hose. To address this, turn off the power switch and let the autoclave sit for at least one hour to cool down, then pour water into the chamber and close the drain valve. In case of objects being put in a disposable plastic bag, quantity of objects to be put inside disposable plastic bag shall be less than 60% of capacity of the chamber and disp...
  • R
    Robin CaseSep 18, 2025
    Why does drainage not stop from my ALP Laboratory Equipment?
    • N
      Nicholas AllenSep 18, 2025
      If drainage from the ALP Laboratory Equipment doesn't stop, it suggests a breakage of the drain valve, possibly due to glass or other objects damaging the valve's Teflon ball. Contact your local distributor.
  • J
    jenniferfarrellSep 22, 2025
    What should I do if the chamber temperature of my ALP CL- 40L Laboratory Equipment doesn't rise?
    • C
      Chad BowenSep 22, 2025
      If the chamber temperature of the ALP Laboratory Equipment does not rise after pressing the START KEY, it could be due to breakage of the heating element or burning of the wire connected to the heating element. Contact your local distributor.
  • H
    hopkinssamuelSep 8, 2025
    What does error code Er3 mean on my ALP Laboratory Equipment?
    • M
      Matthew ShortSep 8, 2025
      If 'Er3' is shown on the ALP Laboratory Equipment display with an alarm, it indicates that the chamber temperature rose 2? higher than the setting temperature during sterilization. Ensure the distance between the chamber wall and objects is more than 10mm. Oleaginous objects must not be sterilized and if the problem persists, contact your local distributor.
  • E
    Eduardo VelasquezSep 5, 2025
    Why is there steam leaking from the lid of my ALP CL- 40L Laboratory Equipment?
    • Z
      zhicksSep 6, 2025
      If there's steam leaking from the lid of the ALP Laboratory Equipment, it could be due to deterioration of the lid gasket (cracks or hardening) or misalignment of the lid and chamber. Contact your local distributor to resolve this issue.

Summary

Safety Information

Safety Definitions and Warnings

Explains hazard symbols, prohibited actions, and strict observations.

Safe Usage and Installation Precautions

Guidelines for operating and installing the autoclave safely.

Operational Safety Guidelines

Specific safety advice during autoclave operation.

Installation and Setup

Power Cable and Grounding

Details electrical connection and safety grounding procedures.

Operation Guide

Basic Operation Workflow

Provides a step-by-step flowchart for basic autoclave use.

Pre-Operation and Setup

Covers power on, drain valve, water supply, and exhaust bottle preparation.

Setting Operation Parameters

Details on setting patterns, temperature, time, and exhaust.

Starting and Monitoring the Cycle

How to start, what to observe during operation.

Completing and Safely Unloading

Steps for cycle completion, cooling, and unloading.

Safety Devices

Overview of Safety Mechanisms

Describes the purpose of safety valve, over pressure detector, and water lack preventer.

Troubleshooting Guide

Maintenance

Spare Parts and Replacement

Technical Specifications