4
How to use autoclave
Installation
Do not use power strip. (Refer to 4.4 Power cable connection and protective grounding)
If power cable is bundled or tangled and /or wrongly connected, these may cause fire and / or get a shock.
Protective grounding conductor must be connected to protective grounding terminal.
Connection to gas pipe or water pipe may cause explosion, get a shock and /or breakage.
Do not install autoclave in the place where chemicals such as flammable materials, explosive
materials, and /or corrosive gas are kept.
Installing autoclave in above mentioned place may cause serious accident and/or trouble of autoclave.
Cleaning
Do not use cleaning products for cleaning the inside of the chamber.
It may cause explosion due to corroded metallic parts.
This autoclave sterilizes objects with high temperature and high pressure.
Do not operate putting following materials and / or dangerous substances
(Refer to 11.3 Dangerous substances list) in chamber.
Explosive materialsFlammable materialsAcid or Alkaline materials
Corrosive materialsNon heat-resistance materials
In case of sterilization of saline objects such as seawater, discharge saline water after
sterilization. Then put fresh water and operate several times without load. If you spill saline
objects, be sure to wipe off.
Failure to do them may cause explosion due to corroded chamber and pipes.
Do not disassemble and / or modify autoclave.
Autoclave being pressure vessel, disassemble and / or modification of autoclave may cause accident and / or
breakage.
Do not move autoclave holding lid nor push it hard.
Lid may jolt out of alignment which leads to serious accident and / or trouble.
Do not get autoclave wet.
If electric component get wet, it may be broken, get a shock, and even cause fire.
This autoclave should be used indoors. Do not install it outdoors.
There are many possibilities of malfunction and/or breakage of autoclave if it is used outdoors.
This autoclave must be installed on level and hard floor.
Installation without stability may cause malfunction and /or trouble.
Do not use liquid which corrode metal and/or paint (EX. thinner, hypochlorous acid, chlorine
dioxide, acetyl hydroperoxide) for cleaning the exterior.
Metallic parts may corrode.
Do not splash chlorine chemical and/or cleaning products (EX. hypochlorous acid, chlorine
dioxide, acetyl hydroperoxide) to the exterior during cleaning the floor.
Metallic parts may corrode.