32
Be sure to set automatic exhaust OFF when sterilizing liquid object or glass container.
(Do not press EXHAUST KEY.)
Quick reduction of chamber pressure may cause breakage of glass container or blow-out of hot liquid.
In case of breakage or spill of objects, be sure to clean inside of chamber to use autoclave
safely, or it may decrease life of heating element and/or chamber, then causes corrosion of
chamber and serious accident.
Do not slide the lid lock lever by force nor open lid till TEMPERATURE DISPLAY indicates below
80℃.
Lid lock can not be released. Because it is mechanically-locked. Forcing the lever to move may damage to
mechanical locking device, it may cause unexpected accident.
Manual exhaust under 121℃ is desirable.
Exhausted steam from the chamber is not cool enough which may cause burn injury.
And water in exhaust cooling tank reduces faster therefore frequent water supply is required.
Be sure to press STOP KEY to release lid lock after operation complete.
If you don’t press STOP KEY, lid lock is not released.
Do not slide the lid lock lever to the right side when you release the lid lock.
The lid lock may not be released which leads to Er4.
WARNING
CAUTION