5.7.3 Exhaust setting
① Set automatic exhaust / referring to following statements.
② In case of , set automatic exhaust starting temperature.
●Automatic exhaust
Automatic exhaust enables to exhaust steam to outside automatically after sterilization to cool
down quickly. It’s necessary to set in each operation depending on objects as follows.
For appliance(except glassware) ・・・
For glassware, liquid or agar medium, other high viscosity substance ・・・
Even in case of , automatic exhaust starting temperature can be set according to boiling
point of objects to prevent bumping. Setting range of automatic exhaust starting temperature
is 102 to 121℃.
Set automatic exhaust or not in each operation depending on object to be sterilized.
The object which is insufferable under rapid pressure and temperature reduction may burst in chamber in case
of automatic exhaust.
●Manual exhaust
Either automatic exhaust , or automatic exhaust after sterilization complete, you can
exhaust manually by pressing EXHAUST KEY before the temperature in the chamber reaches to
automatic exhaust starting temperature.
If you press EXHAUST KEY again, you can stop exhaust.
Manual exhaust under 121℃ is desirable.
Exhausted steam from the chamber is not cool enough which may cause burn injury.
And water in exhaust cooling tank reduces faster therefore frequent water supply is required.