5.10 Unload object (below 80℃ in chamber)
① Check if pressure gauge indicates 0MPa and TEMPERATURE DISPLAY indicates
below 80℃.
② Slide the lid lock lever slowly to open position (right end) with pressing front part of
the lid cover then open lid.
③ Unload objects slowly.
Do not open lid until pressure gauge indicates 0MPa and TEMPERATURE DISPLAY indicates below
80℃.
Steam spouts out of chamber by residual pressure which may cause burn injury.
When sterilizing liquid, it takes longer to cool. Be sure to check the temperature go down
sufficiently then unload the liquid container from chamber slowly, or you may get burned by
bumping of liquid.
Liquid objects keep high temperature after sterilization complete since temperature of liquid cool down slower than
temperature in chamber.
Use a heat-resistance glove when operator unloads objects after operation complete.
If operator touches chamber and/or objects without heat-resistance glove, operator may get burned.
If you spill liquid when unloading objects, be sure to wipe off it.
It may cause malfunction or trouble of autoclave.
5. 11 Close and lock the lid
① Grip lid hold then close lid slowly.
② Slide the lid lock lever slowly to close position (left end) with pressing front part of the
lid cover.
Do not slide the lid lock lever by force.
Forcing the lever to move may damage to mechanical locking device, it may cause unexpected accident.
5.12 Turn off power switch(breaker)
① Drain water in chamber by laying down the lever of drain valve in case autoclave is not
used for a long time.
② Turn off power switch(breaker).
WARNING
CAUTION
WARNING