1. Light Sensor lights LCD screen for 10 seconds in a dark room when any
button is pushed.
2. Test Circuit symbol and an audio warning indicates device is not
treating for one or both channels (the number of the Channel that is not
working will appear). The timer will stop and shut down in 30 minutes if
nothing else is done. Current must be set above 0 µA and moistened
electrodes must be in contact with skin for Alpha-Stim
®
M to work. An
audible signal and resumption of the timer occurs the instant the integrity
check circuit determines everything is working properly.
3. Battery Charge indicator. Replace battery when only 1 bar remains at
which time the device shall give out an audible low battery warning and
repeat the warning every 10 minutes (unless the mute function is on).