10 | 18

4.2.4 Connecting in 6 wire technology
Notice: When connecting in 6 wire technology the jumpers J1 and J2 must be removed. Otherwise the
6-wire technology does not have the desired effect.
Pic. 4:
Connection in 6-wire technology

4.3 Analogue output
The following standardized analogue outputs, depending on the ordered option, are available:
Version 010:
The output is: 0 … +10 Volts (max 1 mA)
Version B10:
The output is: ± 10 Volts (max 1 mA)
Version 420:
The output is: 4 … 20 mA (max 500 ohms)
Other analogue outputs available on request. For parameter settings, further information concerning
scaling, or customized analogue outputs, please refer to the additional sheet “Allocation / Device
Configuration).
SG-Amplifier
SG-Signal
+Sense
+SG-Excitation
-Sense
-SG-Excitation
Transducer cable
350 Ohm
R
wire=0,5 Ohm
R
wire=0,5 Ohm