5.2 Adjustment / calibration of the amplifier
In order to adjust or calibrate the amplifier a multimeter has to be connected to the analogue output.
Inspection:
Unload the measuring device.
Connect the multimeter (refer to chapter 4 “Terminal Assignment”).
Set the analogue output to 0 Volts resp. 4 mA.
Load the measuring device (i.e. the transducer) at least 3 times with its maximum load.
Load the device with 80 % of its maximum load.
Check if the analogue outputs comply with the load and is in the designated specs.
If not, the measurement system might need to be recalibrated. Above that, the installation
position and the overall setup might need an inspection.
Adjustment / Calibration:
The coarse amplification, and thus the voltage output, is determined by an internal resistor (RG) and
can be fine adjusted by a potentiometer (P-02). To adjust the current output, which follows the voltage
output, the potentiometer P-03 is to be used. (See next chapter.)
Before the calibration a warm-up time of approx. 30 mins should be kept. The measuring device has to
be unloaded.
Unloaded means with:
Force transducers: no applied force at all
Pressure transducers no pressure, except atmospheric influence
In this unloaded state adjust the zero-point with the potentiometer P-01.
Apply a defined load. (by calibrated weights, pressure generator etc.)
Adjust the analogue output corresponding to the applied load.
In order to reach the required accuracy, it may be necessary to repeat these steps.