Guidance and manufacturer’s declaration – electromagnetic emissions
Emissions test Compliance
RF emissions CISPR 11 Group 1
RF emissions CISPR 11 Class B
Harmonic emissions IEC 61000-3-2 Class A
Voltageuctuations/ickeremissions Complies
IEC 61000-3-3
Electromagnetic environment – guidance
The Ameda Platinum Breast Pump uses RF energy
only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause
anyinterferenceinnearbyelectronicequipment.
The Ameda Platinum Breast Pump is suitable
for use in all establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes.
The Ameda Platinum Breast Pump is intended for use in
the electromagnetic environment specified below.
The customer or the user of the Ameda Platinum Breast
Pump should assure that it is used in such an environment.
Operational Conditions: Pump
Temperature+53°F(+12°C)to+93°F(+34°C)
RelativeHumidity20%to70%
Atmospheric Pressure 0.82atm(83.1kPa)to1.05atm
(106.4kPa)
Transport and Storage Conditions: Pump
Temperature-10°F(-23°C)to+130°F(+54°C)
RelativeHumidity20%to90%
Atmospheric Pressure 0.82atm(83.1kPa)to1.05atm
(106.4kPa)
Weight: Approximately9.75lbs.(approximately4.42kg)
Dimensions: Length9.75in.(24.7cm)/Width 10.5 in.
(26.7cm)/Height8.0in.(20.3cm)
Standards:
E325189 MEDICAL EQUIPMENT
WITHRESPECTTOELECTRICSHOCK,
FIREANDMECHANICALHAZARDSONLY
INACCORDANCEWITH
IEC60601-1: 1988 + A1:1991 + A2:1995,
IEC60601-1-2,CAN/CSAC22.2No.601.1-M90
MEDICAL ELECTRICAL EQUIPMENT needs special precautions
regarding EMC and needs to be installed and put into service
according to the EMC provided in these Instructions for use.
PortableandmobileRFcommunicationsequipmentcan
affect MEDICAL ELECTRICAL EQUIPMENT.
Warning:TheuseofACCESSORIES,transducersand
cables other than those specified, with the exception of
transducers and cabled sold by the MANUFACTURER of the
Guidance and manufacturer’s declaration – electromagnetic immunity
NOTE:
U
t is the a.c. mains voltage prior to application of the test level.
Immunity test
Electrostatic
discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burstIEC
61000-4-4
Surge IEC 61000-4-5
Voltage dips,
short
interruptions
and
voltage variations on
power supply input
lines IEC 61000-4-11
Powerfrequency
(50/60Hz)magnetic
field IEC 61000-4-8
IEC 60601 test level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines/±1kVfor
input/outputlines
±1kVline(s)toline(s)
±2kVline(s)toearth
<5%
U
t(>95% dip in
U
t)for0,5cycle
40%
U
t(60%dipin
U
t)for5cycles
70%
U
t(30%dipin
U
t)for25cycles
<5%
U
t(>95% dip in
U
t)for5s
3A/m
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines/±1kVfor
input/outputlines
±1kVline(s)toline(s)
±2kVline(s)toearth
<5%
U
t(>95% dip in
U
t)for0,5cycle
40%
U
t(60%dipin
U
t)for5cycles
70%
U
t(30%dipin
U
t)for25cycles
<5%
U
t(>95%dipin
U
t)for5s
3A/m
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relativehumidityshouldbeatleast30%.
Mainspowerqualityshouldbethatofatypical
commercial or hospital environment.
Mainspowerqualityshouldbethatofatypical
commercial or hospital environment.
Mainspowerqualityshouldbethatofatypical
commercial or hospital environment. If the user
oftheAmedaPlatinumBreastPumprequires
continued operation during power mains
interruptions, it is recommended that the Ameda
Platinum Breast Pump be powered from an
uninterruptible power supply.
Powerfrequencymagneticeldsshouldbeat
levels characteristic of a typical location in a
typical commercial or hospital environment.
The Ameda Platinum Breast Pump is intended for use in
the electromagnetic environment specified below.
The customer or the user of the Ameda Platinum Breast
Pump should assure that it is used in such an environment.
Ameda Platinum Breast Pump as replacement parts for
internalcomponents,mayresultinincreasedEMISSIONSor
decreasedIMMUNITYoftheAmedaPlatinumBreastPump.
Warning: The Ameda Platinum Breast Pump should not
beusedadjacenttoorstackedwithotherequipment.If
adjacentorstackeduseisnecessary,theAmedaPlatinum
Breast Pump should be observed to verify normal operation
in the configuration in which it will be used.