ENGLISH 7
Using Timer
•Timerstartswhenpump
is on. Timer displays
pumping duration in
minutes : seconds.
•Timercanberesetby
pressing Timer Reset
button.
•Timerandpump
automaticallyturnOFF
after 60 minutes.
Turning Pump OFF
•Whennishedpumping,
press and hold Power
button to turn pump and
timer functions OFF.
Removing Ameda HygieniKit Milk Collection System
from Breast(s)
•AfterpumpisOFF,ifnecessary,insertangerbetween
breast(s)andbreastanges(s)tobreaksuction.
•Removebreastanges(s)frombreast(s).
Cleaning the Ameda Platinum Breast Pump
For At-Home/End-Users
After each pumping, wipe the Ameda Platinum Breast Pump
with clean, damp cloth.
For Hospitals/Rental Stations
Clean between users. Wearing disposable gloves, wipe
the Ameda Platinum Breast Pump with damp disposable
cloth using hospital approved antibacterial cleaner, such as
CaviCide
TM
.
Troubleshooting
Pump does not turn ON
•Ensureproperpowercordconnectiontopump.
•Ensuregroundedplugconnectiontoelectricaloutlet.
•Ensureoutlethaspower.
Unable to adjust speed
•EnsurepumpfunctionisON.
•Ensure“CheckKit”isnotshowninsuctiondisplay
window.
Unable to adjust suction
•EnsurepumpfunctionisON.
•Ensure“CheckKit”isnotshowninsuctiondisplay
window.
Low or no suction
•Ensure“CheckKit”isnotshowninsuctiondisplay
window.
•Checkbreastangeforadequatesealagainstbreast.
•CheckthattheAmedaHygieniKitMilkCollectionSystem
is properly assembled.
•Checkthatadaptercapisrmlysnappedontotopof
breast flange.
•Ensuresiliconediaphragmisinplaceinsidetopofbreast
flange.
•Ensuretubingadapterisproperlyttedintotubing
adapter port on pump.
•Inspectvalveforstretchingortears.Ifseen,replace
valve.
•Ensurevalveispushedrmlyintoplaceinlowerportion
of breast flange.
•Ensureonesideoftubingadapterisclosedwhensingle
pumping.
“Check Kit” error message appears
•Ensuretubingadapterisproperlyttedintotubingadapter
port on pump.
•Ensureonesideoftubingadapterisclosedwhensingle
pumping.
•Checkthatadaptercapisrmlysnappedontotopof
breast flange.
•Ensuresiliconediaphragmisinplaceinsidetopofbreast
flange.
•Checktubingandsiliconediaphragmtoensurethatparts
are free of defects.
•Checktubingisrmlyattachedtoadaptercap.
“Service Required” error message appears
•Callyourlocaldistributor.Forlistingofdistributors
worldwide, visit www.ameda.com.
The Ameda Platinum Breast Pump should be serviced only
by Ameda, Inc. or authorized Service Centers approved by
Ameda, Inc.
Maintenance
It is recommended the Ameda Platinum Breast Pump be
inspected at least once every year to include the following:
•Visuallycheckpumphousingforcracks,breakage,or
damage and ensure Pump Feet are present.
•Visuallycheckforpowercorddamage.
•Checkforproperoperationofcontrolsandbuttons.
•Checktherearenomissingscrews.
•CheckcoverispresentonServicePortDoor.
•CheckTubingAdapterPortisfreefromobstruction.
Quick Functional Check
Thefollowingchecks/testsshouldbeperformedpriortouse
by a new user:
•Visualchecksforpowercorddamage.
•Checkforabnormalnoise.
•Checkforproperoperationofcontrolsandbuttons.
Service Port
•DonotopenServicePortDooronbottomofthePump.
•Donotconnectanyequipmenttotheserviceportwhen
operating the pump.
•ServiceportistobeaccessedonlybyauthorizedService
Centers approved by Ameda, Inc.
Replacing Detachable Power Cord:
•Powercordisdetachablefrompowerinlet.
•Ifpowercordconditionisindoubt,onlyreplacewith
Amedasupplied3-conductorcordequippedwith
protective ground conductor.
Technical Information
Protection Class 1, Type B
Power Supply:
OperatingInputVoltage100VACto240VAC
OperatingFrequency50/60Hz
Input Power 80VA
FuseRatingQuickActing(F),2.5A,250V,5x20mm
Power Cord: Use only Ameda supplied power cord
For North America, use Ameda 24502059
Minimum Voltage Rating: 125V
Minimum Current Rating: 10A
PlugType:IEC60320C13(HospitalGrade)
MaximumLength:10feet(3.05m)
Cord Type: 3x18AWG, rated VW-1, 105C
Jacket Type: SJT or SJTW
Certication:ULListed(ELBZ)
Operation: Continuous
ISO10079-1,IntermittentSuction,MediumVacuum
Suction: Showninpercentageoffullscalesuction1%-100%;
(approximately30mmHg to 250mmHg/4.0kPa to 33.3kPa)
Speed: 30to80cyclesperminute(cpm)
suction
timer
sucti
o
speed