The effect of electrolyzed reformed water can be easily determined. The reformation of
water is constant and independent of flow within the instrument range. Generally, the
error resulting from air or oxygen streams can be as high as 10 percent. Use the following
technique to reduce this effect:
1. Adjust sample flow to 100 mL/min and note the moisture content.
2. Readjust sample flow to 200 mL/min and note the moisture content.
Moisture Value = Reading at 200 mL/min - reading at 100 mL/min