123199 04 15
3.9 MENU
The MENU function (cf. fig. 6) is used to modify the SETUP
parameters.
Hold down for approx. 2 seconds. The word will
appear on the display.
Press . The first SETUP parameter will be displayed.
Press or
: Temperature unit C or F.
: The highest permissible temperature for
the calibrator.
: Temperature change per minute used
in connection with SWITCH TEST.
: Extra time which must elapse once the
well is stable before the check mark
symbol is displayed.
: Temperature resolution of 0 or 1
decimal.
Press to select the SETUP parameter you wish to
change. The current value will flash.
Press or to select the required value.
Press to accept your selection or to cancel and
return to the previous value.
Once you have changed all SETUP parameters as required,
cancel the function by pressing twice.