18 123199 04
5.0 After use
Warning
Never leave hot insertion tubes which have been removed
from the calibrator unsupervised they may constitute a fire
hazard or personal injury.
If you intend to store the calibrator in the optional carrying
case after use, you must ensure that the instrument has
cooled to a temperature below 100°C/212°F before placing it
in the carrying case.
Caution…
The insertion tube must always be removed from the
calibrator after use.
The humidity in the air may cause corrosion oxidation on
the insertion tube inside the instrument. There is a risk
that the insertion tube may get stuck if this is allowed to
happen.
If the calibrator is to be transported, the insertion tube
must be removed from the well to avoid damage to the
instrument. If the insertion tube is not removed from the
CTC-1200 A the ceramic well might crack.
Scratches and other damage to the insertion tubes
should be avoided by storing the insertion tubes carefully
when not in use.
Caution Hot surface
Do not touch the grid plate, the well or the insertion
tube - they may be very hot and cause burns.
Do not touch the tip of the sensor when it is removed
from the insertion tube/well it may be very hot and
cause burns.
Do not remove the insertion tube from the calibrator
before the insertion tube has cooled down to less than
50°C/122°F.