5.1 Switching off the calibrator
The following routine must be observed before the insertion tube is
removed and the instrument switched off (cf. fig. 1):
1. If the calibrator has been heated up to temperatures above
50°C/122°F, you must wait until the instrument reaches a
temperature below 50°C/122°F before you switch it off.
2. If the calibrator has reached a temperature below 0°C/32°F,
it should be heated momentarily to a temperature of
100°C/212°F.
3. Switch off the calibrator using the power control switch (pos.
7).
4. Remove the insertion tube from the calibrator using the tool
supplied with the instrument.
5. Optional: Store the calibrator in the carrying case.