EasyManuals Logo

Ametek JOFRA ATC-125 A User Manual

Ametek JOFRA ATC-125 A
124 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #113 background imageLoading...
Page #113 background image
105446 06 23-03-2009 113
THERMAL SPECIFICATIONS ATC-650 A/B
Well specifications 40 mm (1.57 inch) axial homogeneity :
0.30°C (0.54°F) @650°C (1202°F)
0.25°C (0.45°F) @320°C (608°F)
0.10°C (0.18°F) @100°C (212°F)
0.05°C (0.09°F) @50°C (122°F)
50 mm (1.97 inch) axial homogeneity :
0.60°C (1.08°F) @650°C (1202°F)
0.35°C (0.63°F) @320°C (608°F)
0.15°C (0.27°F) @100°C (122°F)
0.05°C (0.09°F) @50°C (122°F)
60 mm (2.36 inch) axial homogeneity :
1.00°C (1.80°F9 @650°C (1202°F)
0.40°C (0.72°F) @320°C (608°F)
0.10°C (0.18°F) @100°C (212°F)
0.10°C (0.18°F) @50°C (122°F)
70 mm (2.76 inch) axial homogeneity :
1.40°C (2.52°F) @650°C (1202°F)
0.70°C (1.26°F) @320°C (608°F)
0.25°C (0.45°F) @100°C (2!2°F)
0.10°C (0.18°F) @100°C (122°F)
80 mm (3.15 inch)axial homogeneity :
2.00°C (3.60°F) @650°C (1202°F)
0.90°C (1.62°F) @320°C (608°F)
0.35°C (0.63°F) @100°C (212°F)
0.15°C (0.27°F) @50°C (122°F)
Difference between borings :
0.01°C (0.02°F) @100°C (212°F)
0.02°C (0.04°F) @320°C (608°F)
0.05°C (0.09°F) @650°C (1202°F)

Table of Contents

Other manuals for Ametek JOFRA ATC-125 A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek JOFRA ATC-125 A and is the answer not in the manual?

Ametek JOFRA ATC-125 A Specifications

General IconGeneral
BrandAmetek
ModelJOFRA ATC-125 A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals