120 23-03-2009 105446 06
INPUT SPECIFICATIONS
ACCURACY IN C/F
Accuracy RTD Cu100
(excluding sensor accuracy)
±0.047°C(±0.085°F) @ -90°C(-130°F)
±0.050°C(±0.090°F) @ -50°C(-58°F)
±0.052°C(±0.094°F) @ 0°C(32°F)
±0.060°C(±0.108°F) @ 150°C(302°F)
Accuracy RTD Cu50
(excluding sensor accuracy)
±0.087°C(±0.157°F) @ -90°C(-130°F)
±0.090°C(±0.162°F) @ -50°C(-58°F)
±0.093°C(±0.167°F) @ 0°C(32°F)
±0.100°C(±0.180°F) @ 150°C(302°F)
Accuracy Pt100 reference input
(excluding sensor accuracy)
±0.019°C(±0.034°F) @ -90°C(-130°F)
±0.020°C(±0.036°F) @ -50°C(-58°F)
±0.021°C(±0.038°F) @ 0°C(32°F)
±0.023°C(±0.041°F) @ 155°C(311°F)
±0.024°C(±0.043°F) @ 225°C(437°F)
±0.026°C(±0.047°F) @ 320°C(608°F)
±0.028°C(±0.050°F) @ 425°C(797°F)
±0.032°C(±0.058°F) @ 650°C(1202°F)
±0.034°C(±0.061°F) @ 700°C(1292°F)