A
Amber McguireAug 20, 2025
How to fix no light in display error in Ametek JOFRA ATC-125 Measuring Instruments?
- PpinedajohnAug 20, 2025
If the Ametek Measuring Instruments display shows no light, begin by inspecting the fuses in the mains inlet. Then, check the fuses located on the Power PCB: F1 – F6 for ATC-125, F1 – F7 for ATC-140/155/156/157, and F1 – F7 for ATC-250/320/650. Also, check the supplies and replace the Power PCB if necessary, as a defective regulator on the Power PCB could be the cause.