38 23-03-2009 105446 06
F3F2 F4 F5F1
Press
to continue the calibration without editing the basic
parameters
or,
Press
to start the editor.
F3F2 F4 F5F1
Make the necessary changes, exit the editor by pressing
and continue the calibration by pressing
.
If the sensor under test is a thermocouple sensor and the
manual compensation mode is selected in work orders, a cold
junction temperature must be defined.