105446 06 23-03-2009 99
THERMAL SPECIFICATIONS –
LIQUID BATHS
ATC-140 A/B
Well specifications 40 mm / 1.57 inch axial homogeneity :
0.05°C / 0.09°F @140°C / 284°F
0.05°C / 0.09°F @-20°C / 4°F
50 mm / 1.97 inch axial homogeneity :
0.05°C / 0.09°F @140°C / 284°F
0.05°C /0.09°F @-20°C / 4°F
60 mm / 2.36 inch axial homogeneity :
0.10°C / 0.18°F @140°C / 284°F
0.05°C / 0.09°F @-20°C / 4°F
70 mm / 2.76 inch axial homogeneity :
0.10°C / 0.18°F @140°C / 284°F
0.05°C / 0.09°F @-20°C / 4°F
80 mm / 3.15 inch axial homogeneity :
0.15°C / 0.27°F @140°C / 284°F
0.10°C / 0.18°F @-20°C / 4°F
Radial homogeneity :
0.025°C / 0.045°F
Influence from load :
0.10°C / 0.18°F @140°C / 284°F
0.03°C / 0.05°F @-20°C / -4°F
Influence from load with
Ext. Reference :
0.02°C / 0.04°F @140°C / 284°F
0.02°C / 0.04°F @-20°C / -4°F
Difference between insert / oil:
0.20°C / 0.36°F
Long term drift (1 year) :
±0.05°C / ±0.09°F
Calibration accuracy (test limit) ±0.11°C / ±0.20°F
Temperature coefficient 5ppm/°C (0-40°C) / (32-104°F)
Stability ±0.02°C / ±0.04°F