122770 06 18-09-2007 15
V must have a value between 32.000 and 33.500, when the inputs
are short-circuited. At 10V the value has to be between 56.000 and
59.000.
)
Press in the Service Input Menu and the AD-converter
Status Menu will appear.
Information on A/D-converter status.
2.1.1.3 Service Output Menu
)
Press in the Service Main Menu and the Output Menu will
appear.
Here it is possible to test whether the beeper works and whether the
Sync. relay can change.