EasyManua.ls Logo

Ametek JOFRA CTC-1200 A - 2 Maintenance; Adjusting and Calibrating the Instrument

Ametek JOFRA CTC-1200 A
54 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
123200 03 26-03-2004 7
2.0 Maintenance
2.1 Adjusting and calibrating the instrument
You are advised to return the calibrator to Ametek Denmark A/S or an
accredited laboratory at least once a year for calibration and
adjustment.
Alternatively, you can calibrate/adjust the calibrator yourself.
You will need a calibrated, traceable sensor and thermometer with an
accuracy better than ±0.1°C / ±0.18°F (all MTC and CTC instruments
except CTC-1200 A), ±1.4°C / ±2.5°F (CTC-1200 A only) . Ensure
that the ambient temperature is 23°C±3°C / 73.4°F±5.4°F.
To calibrate/adjust the instrument five different temperatures are
needed. These are all preset in the various types of instruments.
The calibration temperatures are:
CTC-140 A / 1. -15°C / 5°F
MTC-140 A 2. 20°C / 68°F
3. 60°C / 140°F
4. 100°C / 212°F
5. 140°C / 284°F
CTC-320 A/B / 1. 50°C / 122°F
MTC-320 A 2. 120°C / 248°F
3. 180°C / 356°F
4. 250°C / 482°F
5. 320°C / 608°F
CTC-650 A/B / 1. 50°C / 122°F
MTC-650 A 2. 200°C / 392°F
3. 350°C / 662°F
4. 500°C / 932°F
5. 650°C / 1202°F

Other manuals for Ametek JOFRA CTC-1200 A

Related product manuals