Figure 5-140. RUN Menu
The RUN menu has the following fields:
Entry Description
Continuous Causes the transient execution to continue indefinitely. The
execution must be stopped manually.
X Times Determines the number of times a transient list is repeated. The
default value is zero, which means the programmed list runs only
once. The range for this field is from 0 through 99999. This
repeatable function should not be confused with the REPEAT
function available for individual events. The event-specific repeat
value will cause only that event to be repeated, not the entire list.
Start Starts a transient execution. The output relay must be closed, or
an error message will appear, and the transient will not start.
Abort Once the START command has been set, the START
selection-button will change to an ABORT button, which could be
used to stop the run and abort the transient list.
5.8.7 Configuration Screen
The CONFIGURATION screen provides a setup of the output mode of operation, power-on states,
operation profiles, parameter limits, and selection of clock mode.