Sequoia / Tahoe Series User Manual California Instruments
232 M447352-01 REV-A
RS232 Configures the RS-232C communications interface; These
settings must match those set for the communications port of the
user's external controller. The setup parameters are as follows:
Baud Rate: sets the baud rate to either 9600, 19,200, 38,400,
57600, or 115,200 baud. The default setting is 115,200 baud.
Data: sets the number of data bits to either 7 or 8. The default
setting is 8 bits.
Parity: sets the parity to either Even, E, Odd, O or no parity, N.
The default setting is no parity, N.
Stop Bits: sets the number of stop bits to either 1 or 2 bits. The
default setting is 1 stop bit.
Start Bits: always set to 1.
Terminator for Received Messages: LF (ASCII 13) is
necessary, but CR/LF (ASCII 10 / ASCII 13) would be accepted.
Terminator for Transmitted Messages: LF (ASCII 13).
Flow Control: available hardware handshake RTS/CTS;
utilization is recommended, but not mandatory.
Figure 5-242. CONTROL INTERFACE Menu, RS232 Sub-Menu
GPIB Sets the IEEE-488 address; the default is 1. The address could
be set from 0 through 31, though address 0 is often reserved for
the IEEE-488 external controller.
Figure 5-243. CONTROL INTERFACE Menu, GPIB Sub-Menu