AMI Analyzer Manual •
••
• 16
Specifications and Disclaimer
Specifications:
Standard ranges:
Single range: 0 - 25% (optional: 0 – 1.0%, 0 – 2.5%, 0 – 50%, and 0 – 100%)
Sensitivity: 0.5% of full scale
Repeatability: +/- 0.1% of full scale at constant temperature
Operating temperature: 25°F to 115°F (-4°C to 46°C)
Humidity: < 85%, non-condensing
Operational conditions: Pollution degree 2, Installation category I I.
Drift: +/- 1% of full scale in 4 weeks at constant temperature (dependent on sensor)
Response times: 90% of full scale < 13 seconds
Outputs: 0-2.5V nominal (un-calibrated).
Power requirements: Between 12 and 24 VDC, <10 mA
Dimensions: 2.15 Dia x 2" high (not including fittings or leads).
Weight less than 1 lbs.