42 Amico Beds Corporation
Electromagnetic Compatibility
Table A.2 – Electromagnetic Immunity
The Apollo Bed (B-AM1-400X) is intended for use in the electromagnetic environment specified below. The customer or user of
B-AM1-400X should assure that it is only used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic Environment –
Guidance
Electrostatic Discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast transient/burst
IEC 61000-4-4
± 2 kV for power supply lines
± 1 kV for input/output lines
± 2 kV for power supply lines
± 1 kV for input/output lines
Main power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV differential mode
± 2 kV common mode
Main power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
<5% U
Υ
(>95% dip in U
Υ
)
for 0.5 cycle
40% U
Υ
(60% dip in U
Υ
)
for 5 cycle
70% U
Υ
(30% dip in U
Υ
)
for 25 cycle
<5% U
Υ
(>95% dip in U
Υ
)
for 5 s
<5% U
Υ
(>95% dip in U
Υ
)
for 0.5 cycle
40% U
Υ
(60% dip in U
Υ
)
for 5 cycle
70% U
Υ
(30% dip in U
Υ
)
for 25 cycle
<5% U
Υ
(>95% dip in U
Υ
)
for 5 s
Main power quality should be
that of a typical commercial or
hospital environment. If the user
of the Apollo Bed (B-AM1-400X)
requires continued operation during
main power interruptions, it is
recommended that the Apollo Bed
(B-AM1-400X) be powered from an
uninterrupted power supply or a
battery.
Magnetic field at power
frequency
IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical location
in a typical commercial or hospital
environment.
NOTE: U
Υ
is the A.C. main voltage prior to application of the test level.
NOTE: This medical device is a patient support and handling device with an electro-mechanical system, able to achieve
various heights and positions for the litter sections as required per the end user. The device meets this essential
performance at the above compliance levels.