Amoul T6 - Abnormal Power Failure Alarms

Amoul T6
126 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
90
low
low
minute ventilation patient
Check if the set lower
limit is reasonable
Embarrassme
nt
Embarrassme
nt
The respiratory
distress time
exceeds the set
time value
Check condition of the
patient
Check that the set time
value is reasonable
Airway
pressure is
high
Airway
pressure is
high
The set upper limit
is exceeded
Check condition of the
patient
Airway obstruction
Check condition of the
patient
Respiratory hose
is misplaced
Put the respiratory hose
in place
Pmax is set too
low
Correct Pmax
Respiratory hose
is twisted
Check position of patient
and move to an
appropriate position if
necessary
10.3 System alarms
Messages
Alarm
Causes
Remedies
The gas
source
pressure is
less than
2.5bar
The gas
source
pressure is
less than 2.5
bar
The cylinder is not
opened, or the gas in
the cylinder runs out
Open the oxygen
cylinder or replace with a
full cylinder
The cylinder is not
properly connected
Check the connections
and connect the
unconnected parts
The compressed gas
source is defective
Replace with a gas
source which is in good
conditions
Gas supply Tidy the ventilator gas