D-0115687-C (OM008) ā 170 Operating Manual
The output from mains adapter is fitted with electronic circuit protection. In
case of overload the adapter will shut down and the indicator will be
extinguished. When the fault is cleared the adapter will operate as normal.
The input to the mains adapter is protected with a non-replaceable fuse. If
this fails the adapter will not operate.
The mains adapter is the mains disconnect device and therefore the
audiometer should be positioned such that easy access to the mains
adapter is possible.
2.4 Audiometer connections
All the relevant accessory terminals and connections are labelled to
ensure correct identification and connection as follows:-