12
D-0115687-C (OM008) ā 170 Operating Manual
3.9 Data transfer to Audibase or AudiView
Test results stored within the audiometer may be transferred to the Amplivox
Audibase database which is available as an option and must be installed on to a
computer (see Section 12 for the part number). Alternatively, Amplivox
AudiView allows data to be transferred to a computer and subsequently viewed,
annotated & printed. This software is supplied on a CD which includes this
operating manual.
Refer to the installation & operating instructions provided with Audibase or
AudiView for further details.
4 Suggested Sequence of Operation and Test Procedure
The following applies to air conduction measurements. For illustrative
purposes 5dB steps are used. Refer also to ISO 8253 for guidance.
4.1 Audiometry preparation and ambient conditions
Refer to the appropriate audiometric standards and other relevant
publications for guidance on audiometric testing.
Audiometric testing should always be performed in quiet conditions (e.g. a
quiet room or an acoustic booth). The optional Audiocups can provide an
additional level of isolation from ambient noise. For further explanation on
permissible ambient noise levels, please refer to the standard ISO6189.