EasyManua.ls Logo

AmpliVox 260 - A2.4 Free-field Speech Calibration; A2.5 Free-field Warble Tones Calibration

AmpliVox 260
43 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
APPENDIX 2 - FREE FIELD CALIBRATION PROCEDURE
35 MODEL 260 INSTRUCTION FOR USE
It is possible that, because of the characteristics of the listening room or test set-up, the calibration levels above cannot
be achieved because the limit of adjustment is reached for one or more frequencies. Re-arrangement of the listening
room may improve the situation, but if not, the following is a possible solution:
Set all of the frequencies for which calibration can be achieved
For frequencies where this is not possible, adjust each to be a multiple of 5dBs from the required level
Produce a correction table for each frequency for which calibration could not be achieved to be applied to
the output level of the audiometer while conducting a test to relate the instrument display to actual output
level from the speakers.
A2.5.2 Calibrating the Competing Noise Channel
This part of the calibration procedure may be omitted if Free-field speech masking is not required.
Enter the Free-field calibration mode as described in A2.5.1
Press the SPEECH key and the display will change to indicate the option to adjust the competing noise
calibration level the legend “Sp Mask” is used to indicate this
Without changing the setting on the external amplifier use the SIGNAL  keys to adjust the level of the
competing noise to 90dBSPL as measured by the SLM using dBA settings.
Calibrate each channel, pressing the RIGHT and LEFT keys to switch between channels
If necessary it is possible to switch between speech (competing noise) and warble calibration modes by
pressing the SPEECH key again
To store the levels and leave Free-field calibration mode, press the MENU key
A2.6 FREE-FIELD LIVE SPEECH CALIBRATION
Note: as stated in Appendix 1 of this operating manual, users should be aware that there is a growing body of
professional opinion that Live Voice speech audiometry is generally not recommended. Exceptional skill and
concentration are required to achieve accurate and consistent levels.
Connect a microphone to the MIC1 input on the audiometer
Press SPEECH repeatedly to ensure that ‘MIC’ is displayed in capitals at the bottom left of the display
(indicating that the external microphone is selected)
The input signal is adjusted in 1dB steps with the MASKING  keys
Input signal adjustment should be made to adjust for the operator’s voice to peak at the 0dB point on the
LEVEL dB bar graph
If recorded speech has been calibrated no further action is necessary
If recorded speech has not been calibrated, the volume control of the amplifier should be adjusted so that
the SLM reads 90dBSPL at the listening point with a 70dBHL setting on the instrument; note that this is an
approximate setting only, as it is not possible to produce a true calibration signal in live speech

Other manuals for AmpliVox 260

Related product manuals