IMPORTANT SAFETY INSTRUCTIONS
7 MODEL 260 INSTRUCTION FOR USE
The output from mains adapter is fitted with electronic circuit protection. In case of overload the adapter will shut down
and the indicator will be extinguished. When the fault is cleared the adapter will operate as normal.
The input to the mains adapter is protected with a non-replaceable fuse. If this fails the adapter will not operate.
The mains adapter is the mains disconnect device and therefore the audiometer should be positioned such that easy
access to the mains adapter is possible.
2.4. AUDIOMETER CONNECTIONS
All the relevant accessory terminals and connections are labelled to ensure correct identification and connection as
follows:
The relevant part numbers are indicated in Section 12.
For connected parts marked * only connect the accessories supplied with the instrument or supplied
by Amplivox or an Amplivox distributor. These parts have been tested for use with the Model 260
Diagnostic Audiometer for compliance with the standards IEC 60601-1 and IEC 60601-1-2. The use
of accessories other than those specified may compromise compliance with these standards. For
other sockets see Appendix 4.