APPENDIX 3 - EMC GUIDANCE & MANUFACTURER’S DECLARATION
38 MODEL 260 INSTRUCTION FOR USE
Guidance and manufacturer’s declaration electromagnetic immunity (2)
The Model 260 Audiometer is intended for use in the electromagnetic environment specified below. The customer or user
of the Model 260 Audiometer should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
10 Vrms
150kHz to 80MHz
10 V/m
80MHz to 2.8GHz
10 Vrms
150kHz to 80MHz
10 V/m
80MHz to 2.8GHz
Portable and mobile RF communications
equipment should be used no closer to any
part of the Model 260 Audiometer,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter.
Recommended separation distance
d = 1.2√P
d = 1.2√P 80MHz to 800MHz
d = 2.3√P 800MHz to 2.5GHz
where P is the maximum output power
rating of the transmitter in Watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,
a
should be less than the compliance
level in each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE 1 At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and