Page 8 OM020-9 Model 270 Operating Manual
Select phones: This option is only available if a secondary
headphone is enabled; use Softkeys 1 or 2 to
select the required headphone type
FF equiv speech?: This option is only available if either DD45 or
TDH39 is the selected headphone; if activated,
free-field equivalent levels will output to the
headphone in speech mode (see Appendix 1)
Store on 2 of 3?: Automatically stores a threshold if the responses
made to two out of three test signals are at the
same hearing level
Set freefield levels?: This option provides access to the freefield
calibration function; refer to Appendix 2 for details
External talkover?: Select NO to use the internal microphone and
YES to use the MIC 1 input
Display Counter?: Activates a percentage counter for use with
speech testing (refer to Appendix 1)
Select printer: Use the appropriate softkey to select either the
Able (AP1300) or the Martel (MCP8830) printer or
the Sanibel (MPT-II) printer
Omit 750/1k5/3k/6k?: To exclude these frequencies from those
available for testing press the YES key
Present Time: Select the minimum and maximum times that a
stimulus can sound when the PRESENT key is
pressed; either 0.5sec (min) & 10sec (max), or
1.0 sec (min) & 2.0 sec (max)
FF Speech units: The units displayed for freefield speech may be
switched between dBHL and dBSPL