US/50 User's Guide 17
Ultrasound Modality
11. ULTRASOUND METER: Consists of four scales:
Average Power Watts Refers to the total watts applied to the
patient. Has a range of 0 to 20 watts.
Average Intensity Watts/CM² Refers to the power density or watts
per square centimeter. Has a range of
0 to 5 watts/cm².
Peak Power Watts Refers to output at 120 Hz amplitude
modulated.
Peak Intensity Watts/CM² Refers to the power density or watts per
square centimeter at 120 Hz amplitude
modulated.
12. ULTRASOUND ON INDICATOR LIGHT
13. COMBINATION ON INDICATOR LIGHT
14. INTERRUPT ON INDICATOR LIGHT
15. ULTRASOUND TRANSDUCER
Note: Only Use The Transducer(s) Assigned To This Generator. Transducer
serial number is located on the Transducer and on the Generator Label.
16. ULTRASOUND TRANSDUCER CRADLE
17. TRANSDUCER PROTECTION SWITCH: When the transducer is placed
in the cradle, the ultrasonic energy will automatically shut off AND when
the transducer is removed from the cradle, the ultrasonic energy will resume
IF: the Ultrasound Intensity control is not set at the "0" position, the treatment
period has not been completed and the bell has not sounded.