US/50 User's Guide 31
US/50 General Operation and Application Procedures
12. Select an alternate Output Mode, Pulsation, or Surge, and adjust the Pulse/
Surge Rate control if so desired. The Output Rate indicator light will reflect
the output rate selected.
13. When treatment is completed, the ac power will shut off and a bell will sound.
To continue or extend a patient's treatment time, reset the Power/Timer knob
within ten (10) seconds of shut off. To initiate early shut off, turn the Power/
Timer knob counterclockwise to the Off position. The ac power will shut off
and a bell will sound. Reduce the Stimulator Intensity control for Stimulator
Output to the 0/Reset position enabling the audible "click". Thoroughly clean
the pad(s) with warm water and dispose of the used cloth cover(s).