Questions and Answers

  • Y
    yolanda36Aug 15, 2025
    Why doesn't the chamber temperature rise in Amsco EAGLE 3013 Laboratory Equipment?
    • J
      Jeremy SullivanAug 15, 2025
      If the chamber drain line or strainer is plugged in your Amsco Laboratory Equipment, make sure to clean the line or strainer. This should help the chamber temperature rise to the correct setting.
  • R
    Ryan EstradaAug 18, 2025
    What to do if chamber pressure drops in Amsco Laboratory Equipment?
    • D
      Devin MartinAug 18, 2025
      If you notice the chamber pressure and temperature dropping during a cycle in your Amsco Laboratory Equipment, the chamber drain line or strainer might be plugged. Cleaning the line or strainer should resolve this issue.
  • C
    cody17Aug 21, 2025
    How to troubleshoot no power on Amsco EAGLE 3013?
    • L
      Linda AndrewsAug 22, 2025
      If your Amsco Laboratory Equipment shows no sign of power, first, ensure that the facility power source is active and turned on. Then, check the POWER switch on the equipment itself to make sure it is in the ON position.
  • C
    Chad GreenAug 26, 2025
    Why is there water in Amsco EAGLE 3013 after cycle?
    • K
      Kimberly Caldwell DDSAug 26, 2025
      If you find water in the chamber or door channel at the end of a cycle in your Amsco Laboratory Equipment, there are several possible causes: * The drain may be plugged or clogged, so clean the drain. * The steam separator/strainer may be plugged, so clean it. * The drain line check valve may be defective, so replace it.
  • B
    Bethany BennettAug 28, 2025
    Why are wrapped goods wet in Amsco EAGLE 3013 Laboratory Equipment?
    • A
      Antonio WilliamsAug 28, 2025
      If wrapped goods are wet after completing a cycle, or there is excessive vapor in the chamber when the door is opened on your Amsco Laboratory Equipment, the dry time might be too short for the load type. Try increasing the setting for dry time.
  • A
    Adrian CordovaSep 1, 2025
    What causes high pressure and low temperature in Amsco EAGLE 3013 Laboratory Equipment?
    • M
      Michael FrostSep 1, 2025
      If, during the sterilize phase, the pressure is too high and the temperature is too low in your Amsco Laboratory Equipment, it might be due to air trapped in the chamber because of a clogged strainer or drain. Clean the strainer inside the chamber, and flush and clean the drain line and traps.
  • R
    Regina WellsSep 4, 2025
    What to do if I can't select a cycle on Amsco EAGLE 3013?
    • M
      Morgan ClarkSep 4, 2025
      If you are unable to select a cycle on your Amsco Laboratory Equipment, check if the chamber door(s) are open. Close and lock the door(s). The control will beep three times if the door is open when a cycle is selected.
  • M
    Mark PerezSep 8, 2025
    Why is steam blowing out of Amsco EAGLE 3013?
    • M
      Marisa BoydSep 8, 2025
      If steam blows out around the door during the sterilize phase of your Amsco Laboratory Equipment, the door gasket may be worn or defective. Consider replacing it.
  • H
    Heather DudleySep 12, 2025
    Why isn't the jacket pressure rising in Amsco Laboratory Equipment?
    • K
      Kayla MendozaSep 12, 2025
      If the jacket pressure does not rise to the required level in your Amsco Laboratory Equipment, consider the following: * Ensure the steam supply hand valve is open; if not, open the valve. * There may be a flood condition or the steam supply solenoid valve is closed. * The steam supply may be less than 50 psig (3.52 kg/sq.cm) dynamic; check the supply pressure and increase if necessary.

Summary

Summary of Safety Precautions

General Safety Warnings

Summary of critical safety warnings and cautions for operation and servicing.

Installation Verification

Installation Checklist

Checklist to ensure proper installation and operation of the sterilizer.

Service Adjustments and Special Features

Access Code for Cycle Values

Enables or disables an access code to set or change cycle values.

Routine Maintenance

Daily Maintenance

Procedures for daily cleaning and inspection of the sterilizer chamber and components.

Operator Troubleshooting

Replacement Parts

Spare Parts List (Series 3011, 3021, 3013, 3023)

List of part numbers and recommended spares for specific sterilizer series.

Spare Parts List (Series 3031, 3041, 3051, 3033, 3043, 3053)

List of part numbers and recommended spares for other sterilizer series.

Amsco EAGLE SERIES 3013 Specifications

General IconGeneral
BrandAmsco
ModelEAGLE SERIES 3013
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals