Amsco EAGLE
®
SERIES 3000
Gravity And Vacamatic Sterilizers
(3/26/98) P-129359-452
Rev.10
Incorporating Volumes 1, 2 and 3 in a Single Bound Set

Questions and Answers

  • B
    Bryan WashingtonAug 14, 2025
    Why doesn't the chamber temperature rise in Amsco EAGLE 3033 Laboratory Equipment?
    • L
      Larry OliverAug 14, 2025
      If the chamber drain line or strainer is plugged in your Amsco Laboratory Equipment, make sure to clean the line or strainer. This should help the chamber temperature rise to the correct setting.
  • M
    Melissa Wu DDSAug 17, 2025
    What to do if chamber pressure drops in Amsco Laboratory Equipment?
    • T
      tinamartinAug 17, 2025
      If you notice the chamber pressure and temperature dropping during a cycle in your Amsco Laboratory Equipment, the chamber drain line or strainer might be plugged. Cleaning the line or strainer should resolve this issue.
  • L
    Lisa ObrienAug 21, 2025
    How to troubleshoot no power on Amsco EAGLE 3033 Laboratory Equipment?
    • R
      Rachael MitchellAug 21, 2025
      If your Amsco Laboratory Equipment shows no sign of power, first, ensure that the facility power source is active and turned on. Then, check the POWER switch on the equipment itself to make sure it is in the ON position.
  • M
    Melinda BenderAug 25, 2025
    Why is there water in Amsco Laboratory Equipment after cycle?
    • D
      Daniel LeeAug 25, 2025
      If you find water in the chamber or door channel at the end of a cycle in your Amsco Laboratory Equipment, there are several possible causes: * The drain may be plugged or clogged, so clean the drain. * The steam separator/strainer may be plugged, so clean it. * The drain line check valve may be defective, so replace it.
  • J
    Jeff PerezAug 27, 2025
    Why are wrapped goods wet in Amsco Laboratory Equipment?
    • R
      Regina SextonAug 28, 2025
      If wrapped goods are wet after completing a cycle, or there is excessive vapor in the chamber when the door is opened on your Amsco Laboratory Equipment, the dry time might be too short for the load type. Try increasing the setting for dry time.
  • T
    Tracey PattersonAug 30, 2025
    What causes high pressure and low temperature in Amsco EAGLE 3033?
    • T
      Taylor DonaldsonAug 30, 2025
      If, during the sterilize phase, the pressure is too high and the temperature is too low in your Amsco Laboratory Equipment, it might be due to air trapped in the chamber because of a clogged strainer or drain. Clean the strainer inside the chamber, and flush and clean the drain line and traps.
  • C
    Christopher RamirezSep 3, 2025
    What to do if I can't select a cycle on Amsco Laboratory Equipment?
    • J
      John MooreSep 3, 2025
      If you are unable to select a cycle on your Amsco Laboratory Equipment, check if the chamber door(s) are open. Close and lock the door(s). The control will beep three times if the door is open when a cycle is selected.
  • D
    Dr. Miguel Stewart DVMSep 6, 2025
    Why is steam blowing out of Amsco EAGLE 3033 Laboratory Equipment?
    • S
      Shane CoxSep 6, 2025
      If steam blows out around the door during the sterilize phase of your Amsco Laboratory Equipment, the door gasket may be worn or defective. Consider replacing it.
  • A
    Ashlee DavisSep 10, 2025
    Why isn't the jacket pressure rising in Amsco EAGLE 3033?
    • S
      steven24Sep 10, 2025
      If the jacket pressure does not rise to the required level in your Amsco Laboratory Equipment, consider the following: * Ensure the steam supply hand valve is open; if not, open the valve. * There may be a flood condition or the steam supply solenoid valve is closed. * The steam supply may be less than 50 psig (3.52 kg/sq.cm) dynamic; check the supply pressure and increase if necessary.

Summary

SUMMARY OF SAFETY PRECAUTIONS

WARNING - BURN AND SHOCK HAZARD

Hazards related to burns and electrical shock.

WARNING - SLIPPING HAZARD

Hazards related to slippery floors.

INSTALLATION VERIFICATION

Installation Checklist

Checklist to ensure correct installation and operation.

Filling;Starting Optional Electric Steam Generator Boiler

Steps to fill and start the optional electric steam generator.

SERVICE ADJUSTMENTS AND SPECIAL FEATURES

SERVICE ADJUSTMENTS (Dip Switch)

Table listing service adjustments controllable via dip switches.

SERVICE ADJUSTMENTS (Field Test Mode)

Table listing service adjustments configurable in field test mode.

ROUTINE MAINTENANCE

Daily

Daily cleaning and maintenance procedures for the sterilizer.

Weekly

Weekly cleaning and checks for the sterilizer and steam generator.

OPERATOR TROUBLESHOOTING

OPERATOR TROUBLESHOOTING CHART

Chart listing common problems, their causes, and corrections.

Amsco EAGLE SERIES 3033 Specifications

General IconGeneral
CategoryLaboratory Equipment
TypeSteam Sterilizer
ModelEAGLE SERIES 3033
ManufacturerAmsco
Sub-CategoryAutoclave
Sterilization Temperature Range105°C to 138°C
Power Requirements208-240V, 60 Hz, 3 Phase