15. Verify that the fluid containers have the correct fluids for this procedure.
Verify that the fluid supply containers are configured according to the screen above and press the check mark () button.
The following screen will be displayed.
16. Fill fluid containers.
To ensure that you have enough fluids to run a complete IDF/SDF procedure, you must begin with fluid levels above the
Minimum Level lines on the chemical containers and at least 15 ml of each enzyme. Add fluids and enzymes as
necessary.
Do NOT leave the enzyme ports on the instrument open to the air or the enzymes
in the valves may dry up and plug the ports.
With all fluid containers filled to the proper levels, press the check mark () button on the Touch Screen Display. The
following screen will be displayed.
Nitrogen supply ON?
AOAC
991.43
IDF/SDF
Fluid levels correct?
AOAC
991.43
IDF/SDF
Amylase -> Amylase
AMG -> AMG
HCL -> HCL
Buffer -> Mes Tris Buffer
Check supply containers
AOAC
991.43
IDF/SDF