20
Setup with G0375A and MU195040A ED for jitter tolerance test
32Gbaud Power PAM4 Converter
G0375A
Noise Generator
MU195050A
32G 2ch SI PPG
MU195020A
(Opt-020, 021, 031)
DUT
Signal Quality Analyzer-R
MP1900A
J1741A cable x4
(80 cm, skew <1 ps)
J1728A cable x2 (40 cm, skew <1 ps)
Synthesizer
MU181000B
Jitter Modulation
MU181500B
J1728A cable x2 (40 cm, skew <1 ps)
XData
Data
41KC-6 x4
Note:
Connect xData output to xData1 input of ED for Clock Recovery.
Connect Data output to Data2 input of ED for sequential measurement.
Supports PAM4 Jitter tolerance test using MU195040A PAM4 sequential
measurement
32G 2ch SI ED
MU195040A
(Opt-020, 022)