3.8 Measurement Results
3-41
3
Measurement
[9] Cal Status Displays the Cal Status.
Table 3.8.1.1-2 Cal Status Display List
Cal Status
Display
Apply
Calibration
OK or Uncal On
Off Off
Note:
Even when Apply Calibration is On, NF Calibration data is
disabled if Cal Status becomes Uncal. When Uncal is displayed,
correct measurement results cannot be obtained. Re-execute NF
Calibration.
[10] ENR Status Displays the ENR Status.
Table 3.8.1.1-3 ENR Status Display List
ENR Status ENR Mode
Spot Spot
Table Table
Note:
When ENR Mode is Table, ENR Status may be displayed in red.
This occurs because Meas Table is not set.
Check the setting of Meas Table when ENR Status is displayed in
red.
[11] Lo Freq / IF Freq:
Displays local frequency or IF frequency that is
fixed according to LO Mode when DUT Mode is not
Amplifier.
Table 3.8.1.1-4 Displayed fixed frequency
LO Mode Displayed fixed frequency
Fixed LO Freq: Local Frequency
Variable IF Freq: IF Frequency