5
Sprint eVo
1
/eVo
2
/eVo
3
Quickstart guide
Quickstart Guide
Switch On
Important: Please connect flue probe first before switching on your
Sprint unit, ensure the water trap is empty and is fitted in the right
direction. Do not insert probe in flue until after auto zero. Always start
in clean air. The gas exhaust outlet is not blocked.
Press and hold the ON/OFF/ENTER key for two seconds. Sprint will emit a few
rising beeps, performs a screen test and takes you directly to the Test Menu.
The instrument can now be used for Pressure and
Temperature functions. When entering Flue Gas Analysis
or CO Room Safety functions are selected the pump
switches on to purge any residual gas from the unit and
carry out a zero. On stabilisation of the instrument the
function can now be used.
Auto Zero
When entering Flue Gas Analysis or CO Room Safety options from the Test
Menu, Sprint will perform an Auto Zero. Ensure you are in clean air before
proceeding.
Tip: Auto zero Sprint outside of the building or well away from the heating appliance to avoid any
potential gases in the vicinity affecting the auto zero process.
Ensure the sensor reading is stable then press proceed þkey to begin auto
zero. The instrument will display the menu option choosen after a suc-
cessful auto zero.
Note: Sprint will switch off the pump when not performing tests or purging.
Test Menu
Fuel Options
Units of Measurement
Anayser Settings
Stored Logs
14:44:27
Cal due
Main Menu
08-Feb-2006
17-Jan-2007
Tip: see soft key
icon list on the
inside front cover
of this manual