Statistics:
Allows you to find out the number of bursts recorded, the number of bytes free as well as the number of measurements
which can be recorded.
D.5.1 Synthesizer configuration
D.5.1.1 Configuration of synthesizer points
The configuration of the synthesizer points is identical to that of the default values.
D.5.1.2 Configuration of synthesizer parameters
The configuration of the synthesizer parameters is performed in the configuration/Synthesizer/Parameters menu,
obviously providing the Synthesizer mode has been confirmed.
Warning: The values (points) must be recorded first.
Once you have selected the Configuration/Synthesizer/Parameters menu, confirm by pressing the VAL key.
This screen allows you to configure the emission:
First point:
This is the 1
st
point to be emitted. It is not necessarily the 1
st
point in the table of values.
Last point:
This is the last point to be emitted. It is not necessarily the last point in the table of values but this point number must be
below the number of points recorded.
Should this not be the case, it will be impossible to record the configuration of the synthesizer parameters
Duration:
This is the amount of time required for the emission of all the points that need to be emitted (Last point - first point).
Repetition:
This is the number of cycles that need to be performed.
Time frame:
This is the time lapse between 2 repetitions.