Loading...
Thermocouple Emission function G
NTA47230-300A6 95 / 101
G.2 Thermocouple emission function (TC6621)
Rated maximum voltage in common mode: 60 VDC or VAC.
G.2.1 Constant voltage
The voltage emission is made by configuring the device as follows:
Thermocouple: indifferent.
Unit: mV.
CSF: OFF.
Gauge
Scope of measurement
Resolution (min)
Precision / 1 year
80 mV
-9.5 mV / 80 mV
1 µV
0.020% R + 3 µV
Temperature coefficient < 10 ppm R /°C from 0°C to 18°C and from 28°C to 50 °C.
Internal resistance: 1 .
G.2.2 Temperature per thermocouple
Type of sensors:
- Standardised in accordance with CEI 584-1/1995 (Thermocouples K, T, J, E, S, B, N).
- In accordance with Din 43710 (thermocouples U and L).
- In accordance with the HOSKINS table (thermocouple C).
- In accordance with the ENGELHARD table (platinum thermocouple).
- Standardised in accordance with ASTM E 988-96 /2002 (Couple D W3Re/W25Re)
- Standardised in accordance with ASTM E 1751-00 /2000 (Couple G W/W26Re)