19
TENS
B Cycle Bursts, 2Bursts/sec, 9pulses/Burst, 100Hz, width is adjustable.
C Continuous mode. Pulse rate, pulse width and intensity are full adjustable.
M
Modulated Width. Pulse width is automatically varied in an interval of 6 seconds. The modula-
tion range of pulse width is from setting value to 35% less than the control setting value. And
then returns to the setting value. Rate, width and intensity are fully adjustable.
SDR
Strength-Duration and Rate modulation. SDR consists of alternating modulated intensity (am-
plitude) and pulse rate so that one parameter is always increasing while the other is decreasing
and vice-versa. The stimulation intensity decreases 20% from the intensity control setting and
returns to the setting value; the pulse rate increase 45% from the modulated rate to the setting
value and return to the modulated rate. Total cycle time is 6 seconds. Rate, Width and Intensity
are full adjustable.
SDW
Strength-Duration and Width modulation. SDW consists of Alternating modulated intensity
(amplitude) and pulse width so that one parameter is always increasing while the other is de-
creasing and vice-versa. The stimulation intensity decreases 20% from the intensity control
setting and returns to the setting value; the pulse width increase 54% from the modulated width
to the setting value and return to the modulated width. Total cycle time is 6 seconds. Rate, width
and intensity are full adjustable.