TECHNICAL SPECIFICATIONS
Channel Dual, isolated between channels
Pulse amplitude Adjustable 0-80mA peak into 500 Ω load each channel, constant current
Pulse Rate 1Hz-160Hz (adjustable), 1Hz/step (1-20Hz), 5Hz/step (20-160Hz)
Pulse Width
a. 50-260μs adjustable for TENS, 10μs/step
b. 250μs xed for EMS
Software ramp up
feature
By changing mode, the output will reset to zero then ramp up to its original setting
intensity. It will protect users from a sudden surge
Patient Compliance
counter
Shows the treatment times in hours
Patient Lock System Prevents the user from changing any xed parameters set by the physician or li-
censed practitioner
T
imer 15, 30, 60 minute and continuous mode selectable
LCD TENS - Show modes, pulse rate, pulse width, timer and CH1/CH2 on LCD panel
EMS - Show modes, pulse rate, on time, off time, timer and CH1/CH2 on LCD panel.