22
Declaration – electromagnetic emissions and immunity –
for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
The ME declaration – electromagnetic immunity
The ME system is intended for use in the electromagnetic environment specied below.
The customer or the user of the ME system should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3V
3V/m
Portable and mobile RF communications equipment should be
used no closer to any part of the EQUIPMENT or SYSTEM
including cables, than the recommended separation distance
calculated from the equation applicable to the frequency of the
transmitter.
Interference may occur in the vicinity of
equipment marked with the following symbol.
Declaration – electromagnetic immunity
The ME system is intended for use in the electromagnetic environment specied below.
The customer or the user of the ME system should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic dis-
charge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic tile. If oors are
covered with synthetic material, the relative humidity should
be at least 30 %.