6. Maintenance
6.1 General Information
1.
This unit is oil-less. Do Not Lubricate.
2.
Risk of electric shock. Do not disassemble the main unit.
3.
All maintenance must be performed by qualified service personnel.
6.2 Filter Change
NOTE - Change the filter once every month or when filter turns gray. Please follow the
instructions below as shown to the right.
1. Open the filter cap.
2. Remove dirty filter with a small, pointed object. Discard the filter.
3. Replace with a clean filter. Additional filters should be purchased from
your provider.
4. Put the filter cap back.
CAUTION - Do not use cotton or any other material as a filter. Do not
wash or reuse the filter. Do not operate the unit without a
filter.
7. Troubleshooting
If your Mini Plus Compressor Nebulizer fails to function, consult the Troubleshooting guide below. If
the problem persists, consult your equipment provider.
Problem
Cause and Solution
Device
doesn’t
operate.
Check if plug is properly fitted into an appropriate electrical outlet.
When device has been running
continuously for over 30 minutes right before
using, an auto shut down may be activated by the built-in thermal protector;
cool down device for 30 minutes before next usage.
Weak
Nebulization
Check for proper electrical voltage.
Check tubing for blockage or air leakage at connection to Mini Plus
compressor or nebulizer cup, and replace as needed.
Check the nebulizer cup to see if it is properly assembled and not
damaged.
If there is any damage, replace as needed.
Check if filter is too dirty, replace as needed.