103N / 105N / 106
EN
2-2 Environmental Conditions
2-3 Electrical Specications
(1) DC Volts
Indoor use.
Maximum Altitude : 2000 Meter.
Pollution Degree : 2
Installation Category : EN61010-1,EN61010-2-033, 1000V
Cat.II, 600V Cat.III.
Accuracy is ± (% reading + number of digits) at 23°C ± 5°C ,
less than 80% R.H.
Input Impedance : 10MΩ . (over 1000MΩ in 400mV range).
The circuits not connected to mains.
The circuits directly connected to Low-voltage installation.
The building installation.
The source of the Low-voltage installation.
Application FieldCAT
II
III
IV
I
Range Resolution
Accuracy
103N 105N / 106
±(0.3%reading +
2digits)
400mV 100μV
40V 10mV
4V 1mV
400V 100mV
1000V 1V
±(0.3%reading +
2digits)
±(0.1%reading +
2digits)
±(0.4%reading +
2digits)
±(0.25%reading +
2digits)
Over voltage
protection
1000V rms
(2) AC Volts
Range Resolution
Accuracy
103N 105N / 106
±(2.0%reading +
8digits)*
400mV 0.1mV
40V 10mV
4V 1mV
400V 100mV
750V 1V
±(2.0%reading +
8digits)*
***
±(1.5%reading +
5digits)
40Hz to 60Hz
±(1.0%reading +
5digits)
60Hz to 1KHz
±(1.3%reading +
5digits)**
±(1.5%reading +
5digits)
40Hz to 60Hz
±(1.3%reading +
5digits)
60Hz to 1KHz
Over voltage
protection
1000V rms
Accessories : Protective Holster , battery, instruction
manual, test leads.