EasyManua.ls Logo

APPA 103N - Page 7

APPA 103N
112 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
103N / 105N / 106
EN
2-2 Environmental Conditions
2-3 Electrical Specications
(1) DC Volts
Indoor use.
Maximum Altitude : 2000 Meter.
Pollution Degree : 2
Installation Category : EN61010-1,EN61010-2-033, 1000V
Cat.II, 600V Cat.III.
Accuracy is ± (% reading + number of digits) at 23°C ± 5°C ,
less than 80% R.H.
Input Impedance : 10MΩ . (over 1000MΩ in 400mV range).
The circuits not connected to mains.
The circuits directly connected to Low-voltage installation.
The building installation.
The source of the Low-voltage installation.
Application FieldCAT
II
III
IV
I
Range Resolution
Accuracy
103N 105N / 106
±(0.3%reading +
2digits)
400mV 100μV
40V 10mV
4V 1mV
400V 100mV
1000V 1V
±(0.3%reading +
2digits)
±(0.1%reading +
2digits)
±(0.4%reading +
2digits)
±(0.25%reading +
2digits)
Over voltage
protection
1000V rms
(2) AC Volts
Range Resolution
Accuracy
103N 105N / 106
±(2.0%reading +
8digits)*
400mV 0.1mV
40V 10mV
4V 1mV
400V 100mV
750V 1V
±(2.0%reading +
8digits)*
***
±(1.5%reading +
5digits)
40Hz to 60Hz
±(1.0%reading +
5digits)
60Hz to 1KHz
±(1.3%reading +
5digits)**
±(1.5%reading +
5digits)
40Hz to 60Hz
±(1.3%reading +
5digits)
60Hz to 1KHz
Over voltage
protection
1000V rms
Accessories : Protective Holster , battery, instruction
manual, test leads.
6