(3) DC Current
AC Conversion Type :
103N / 105N : Average sensing rms indication.
106 : AC conversions are ac-coupled, true rms responding,
calibrated to the rms value sine wave input.
The specied accuracy is for since wave at full scale and
non-sine wave at half scale below 500Hz with crest factor
up to 2.
Input impedance : 10MΩ // less than 100PF.
* Frequency Response : 50Hz to 60Hz.
** Frequency Response : 40Hz to 300Hz.
*** Frequency Response : 40Hz to 500Hz for 4V Range.
AC Conversion Type : AC conversions are average
sensing , calibrated to the rms value sine wave input.
Overload Protection :
1A , 600V IR 10KA fuse (Bussmann BBS-1 or equivaknt) for
mA input. (Size 10.3mm x 34.9mm)
15A , 600V IR 100KA fuse (Bussman KTK 15 or equivalent)
for A input. (Size 10.3mm x 38.1mm)
Range Resolution
Accuracy
103N 105N / 106
±(0.6%reading +
2digits)
40mA 10μA 300mV max
10A 10mA 3V max
400mA 0.1mA 3V max
±(0.4%reading +
2digits)
±(0.7%reading +
2digits)
±(1.2%reading +
3digits)
±(0.5%reading +
2digits)
±(1.0%reading +
3digits)
Voltage
Burden
(4) AC Current
Range Resolution
Accuracy
103N 105N / 106
±(1.5%reading +
5digits)
40mA 1μA 300mV max
10A 10mA 3V max
400mA 0.1mA 3V max
±(1.5%reading +
5digits)
±(2.5%reading +
5digits)
±(2.0%reading +
5digits)
Voltage
Burden
Frequency Response : 40Hz to 1KHz.
AC Conversion Type :
103N / 105N : Average sensing rms indication.
106 : AC conversions are ac-coupled, true rms responding,
calibrated to the rms value sine wave input.
The specied accuracy is for since wave at full scale and
non-sine wave at half scale below
500Hz with crest factor up to 2.